Technical Data and References

WaveLight® Allegretto Wave® Eye-Q Laser

Technical Data

Laser class Class 4 (IEC + 21CFR)
Laser type ArF-excimer laser
Wavelength 193 nm
Pulse duration 12 ns
Repetition rate 400 Hz
Treatment time 2 s/D at a fully corrected OZ of 6.5 mm
Aiming beam diodes 635 nm
Laser class 2 (IEC 60825), class II (21 CFR 1040)
Focusing beam diodes 635 nm
Laser class 2 (IEC 60825), class II (21 CFR 1040)
Spot characteristics Diameter 0.95 mm
Gaussian beam profile
Ablation profiles Wavefront Optimized® ablation profiles
Working principle Flying spot laser system
2 mirrors on closed-loop galvo scanners
Eye-tracking system Active IR eye-tracking system
Eye-tracking and laser trigger synchronized
Centered on the optical center of the pupil
Center manually adjustable
Microscope Zeiss Pico*
Magnification steps 5; 7.5; 12.5; 20; 31.25
Working distance 20 cm (7.9 inches)
Patient adjustment Z- and xy-adjustment for exact positioning
of patients' eye
Applications Wavefront Optimized® refractive
treatments
Wavefront-Guided (A-CAT) treatments in
conjunction with the WaveLight® Allegretto
Wave® Eye-Q Laser Device (optional)
Refraction increments 0.25 D (0.01 D)
Calibration Fluence on PMMA and eye-tracking system
on test targets once a day
Energy control Closed-loop energy control
Laser control Membrane key pad, joystick
Gas supply 1 ArF premix gas cylinder (20 l integrated)
1 nitrogen gas cylinder (20 l integrated or
50 l external)
Nitrogen grade 4.0 or higher
Cooling Air cooled, low noise
Weight Laser with standard bed: 444 kg (979
pounds) (excl. gas)
Laser with swiveling bed: 476 kg (1,050
pounds) (excl. gas)
Power Requirements  
Mains connection 208-240 V single phase 50 Hz ± 5%, 16 A
Power consumption Laser console (in operation): 1.0 kW (max
3.2 kW)
Laser console (in standby): 0.3 kW
Patient bed: 0.6 kW
Plume evacuator: 1.2 kW
Operating Conditions  
Space requirements Laser with standard bed (L x W x H): 250 x
150 x 130 cm (98 x 59 x 51 inches)
Laser with swiveling bed (L x W x H): 250 x
220 x 130 cm (98 x 86 x 51 inches)
Static requirements Capacity >500 kg/m2 (1,103 pounds/10.8
sq foot)
Temperature range 18-300C (65-860F)
Humidity 20-70% at 250C (770F), not condensing
Air condition 1.5 kW thermal output to be carried off
Ventilation Ventilation of min 100 m3 (131 yards3)
required
Integrated Accessories Optional Accessories
Patient bed Swiveling patient bed
Laser gas premix ArF Laser safety spectacles
Ablation depth micrometer Uninterrupted power supply (110 V/60 Hz
or 230 V/50 Hz)
TFT eye-tracking monitor Sterile covers
Notebook computer Sterile sets
Advanced microscope
system
Video system for observing and recording
the treatment
External nitrogen
connection
Surgeon's chair with or without armrests
Plume evacuator
Fiber optic slit
projector with light
source (slit lamp)
 
Compliance  
In compliance with Medical Device Directive (MDD) 93/42/EEC,
IEC 60601/IEC 60825
Certified by LGA InterCert notified body no 1275
Electrical safety class Class I, type B
Medical device class Class IIb according MDD
US approval (PMA) by
Food and Drug
Administration (FDA)
PMA no P030008 (hyperopia and mixed
astigmatism)
PMA no P020050 (myopia with and without
astigmatism)

Specifications are subject to change.

*Trademark(s) are the property of their respective owner(s).